E.12 How do I know when a column is dirty?

See the troubleshooting section of the column manual.

E.13 How do I clean a column?

See the troubleshooting section of the column manual.

E.14 Why is the conductivity high?

Possible reasons for high conductivity include:

The suppressor is not on. Turn on the suppressor from the Chromeleon
Control panel or the ICS-2100 touch screen

SUPPRESSOR

page.

The suppressor needs regeneration. See the suppressor manual for
troubleshooting information. The suppressor manual is included on the
Dionex Reference Library CD-ROM (P/N 053891).

The cell is out of calibration. See

Section 5.1.3

for calibration instructions.

See

Section 4.13

for additional troubleshooting information.

E.15 How do I configure and operate the auxiliary valve?

See Installing the ICS-1100/ICS-1600/ICS-2100 Auxiliary Valve (Document No.
065288), provided in the Auxiliary Valve Kit.