5 • Service

Doc. 065291-01 3/09

101

Use the Wellness Panel to perform the following functions:

Calibrate the pump flow rate

Calibrate the vacuum degas assembly

Calibrate the conductivity cell

Test the conductivity cell electronics, using a dummy cell

Upload calibration values from the ICS-2100 to the Chromeleon
Wellness database

After completing a calibration procedure from the touch screen,
always update the Chromeleon Wellness database by clicking the
Upload button on the Chromeleon Wellness Panel (see

Figure 5-1

).

This action ensures that the current calibration information is stored
in Chromeleon as well as in the ICS-2100.