5 • Service

Doc. 065291-01 3/09

113

5.3

Replacing Tubing and Fittings

The ICS-2100 is plumbed with the tubing and tubing assemblies listed below.

10-32 fittings (P/N 043275) and ferrules (P/N 043276) are used for most
tubing connections. For tightening requirements, refer to Installation of
Dionex Liquid Line Fittings
(Document No. 031432). The manual is included
on the Dionex Reference Library CD-ROM (P/N 053891).

1/8-in fittings (P/N 052267) and ferrules (P/N 048949) are used for
connections to the suppressor

REGEN OUT

port and the eluent reservoir.

Tubing Size and Type

Color

Used For

0.125-mm (0.005-in) ID PEEK
(P/N 044221)

Red

Connection from pump pulse damper
to pressure transducer

0.25-mm (0.010-in) ID PEEK
(P/N 042690)

Black

Connections between other system
components

0.50-mm (0.020-in) ID PEEK
(P/N 042855)

Orange

Connection from injection port to
injection valve

0.75-mm (0.030-in) ID PEEK
(P/N 044777)

Green

Connection from injection valve to
waste

1.58-mm (0.062-in) ID Teflon
(P/N 014157)

Clear

Connection from pump to degas or
eluent reservoir; pump waste

25

μL sample loop

(P/N 042857)

Orange

Connection between ports

L (1)

and

L (4)

on the injection valve