Table 5-6. Utilities Drop-down Menu

The Utilities menu selections relate to system options, and remain the same in each displayed screen. The
Utilities menu consists of the following:

Validate Method: Verifies that the devices required by the method Protocol have been defined or
selected in the Setup screen. Validation is automatically run at the start of a new run.

Check Hardware connections: Checks that all devices listed in the Setup screen are electrically
connected to the system. The mixer is always assumed to be connected, so it does not appear in the
Setup screen and its connection is not checked.

System Information: Displays the current system configuration, including each instrument in the
system and its firmware version number, the Windows

®

version number, available hard disk space, and

the number of methods and runs in the database.

Diagnostics Information: Displays information for service diagnostic purposes only.

Gradient Pump Calibration: Available only from the Manual screen, and it is typically used only after
servicing of the pump. It allows the user to calibrate the gradient pump flow rate and zero the system
pressure gauge.
Note: You must exit the Calibration screen in order for the system to accept the calibration values.

Conductivity Flow Cell Constant Calibration: Allows the user to calibrate the conductivity flow cell.
The conductivity flow cell constant is printed on a tag attached to the flow cell cable. When a new
conductivity flow cell is installed, use this value to enter the flow cell constant.
Note: You must exit the Calibration screen in order for the system to accept the calibration values.

pH Probe Calibration: This utility allows the user to calibrate the pH monitor. It is typically used at the
start of each day’s use of the system.
Note: You must exit the Calibration screen in order for the system to accept the calibration values.

EGP Calibration: This is informational only. It reminds the user to calibrate the pump through the pump
software. Refer to the EGP Instruction Manual.

File

View

Options

Window

Help

BioLogic Duo-Flow - <user name> - <project name> - <method name>

Edit

Utilities

Validate Method

Check Hardware connections

System Information

Diagnostics Information

Gradient Pump Calibration

Conductivity Flow Cell Constant Calibration

pH Probe Calibration

EGP Calibration

INTRODUCTION TO THE SYSTEM SOFTWARE