Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 32

Advertising
background image

2.4.2

Maximizer Mixer

The Maximizer mixer provides the larger capacity required when the Maximizer is used. The Maximizer
mixer may be used with or without a mixer barrel: the mixer body, mixer top, and mixer barrel are provided
with the system; an optional mixer barrel extender is available. The mixer barrels fit between the mixer body
and the mixer top and are used for higher flow rates. Refer to the following table to determine the
appropriate mixer volume.

Table 2-11.

Mixer Barrels and Mixer Capacity for the Maximizer Mixer

Flow Rate

Barrel Extension

Capacity

Assembly Screws

0.5 to 10 ml/min

none

750 µl

1/4

-20 x

1/2

” (1.3 cm)

10 to 40 ml/min

5 ml Barrel Extender

5 ml

1/4

-20 x 1

1/2

” (3.8 cm)

40 to 80 ml/min

12 ml Barrel Extender

12 ml

1/4

-20 x 3

1/4

” (8.2 cm)

The procedure for changing the mixer capacity is provided in the following section.

Figure 2-3. Maximizer Mixer and Mixer Barrel Extender

MIXER

TOP

MIXER

BODY

12 ml MIXER

BARREL EXTENDER

OUTLET PORT

INLET PORT

5 ml MIXER

BARREL EXTENDER

MAXIMIZER MIXER

DESCRIPTION OF BIOLOGIC DUOFLOW SYSTEM

SYSTEM OVERVIEW

2-19