0 system plumbing, System plumbing system installation and setup 4-1 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 93

Advertising
background image

4.0

SYSTEM PLUMBING

This chapter discusses recommended plumbing practices and provides general guidelines for system setup.

The system will work more efficiently if tubing lengths are as short as possible. Bio-Rad provides precut and
1/4-28 fitted, labeled tubing in the Fittings Kit. The illustration below shows where the tubing is designed to
be connected. Discussion of how to create your own tubing can be found in the following section.

Use 1/16” (1.6 mm) OD PEEK tubing and 1/4-28 fittings. Use orange PEEK 0.020” (0.51 mm) ID tubing for
the F10 pumps, and green PEEK 0.030” (0.76 mm) ID tubing for the F40 pumps. Premade tubing kits are
provided with your Fittings kit.

Figure 4-1. System Plumbing using Tubing Kit from Fittings Kit

COLUMN

AVR7-3 SAMPLE
INJECT VALVE

UV DETECTOR

CONDUCTIVITY
MONITOR

TO FRACTION
COLLECTOR

A

B

INLET A/B

INLET A/B

PUMP

#1

#5

#6

#4

#3

PUMP

MIXER

QUADTEC DETECTOR

CONDUCTIVITY

MONITOR

#3

WASTE

WASTE

#2

COLUMN

COLUMN

MIXER

LOOP

AVR7-3 INJECT VALVE

7

1

2

3

4

5

INJECT

PORT

6

#2

SYSTEM PLUMBING

SYSTEM INSTALLATION AND SETUP

4-1