Appendix a. specifications – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 220

Advertising
background image

SPECIFICATIONS

APPENDIX A

A-1

APPENDIX A.

SPECIFICATIONS

System Power Requirements: BioLogic DuoFlow Controller, Workstation, and Maximizer

100 V ~ 5.3 A

230 V ~ 3.5 A

50 - 60 Hz

DuoFlow Dell Controller

Pentium IV

®

Processor, 20 GB hard drive, 256 MB RAM

SoundNIC and video 4MB graphics performance accelerator

Dell M781 multimedia 16” monitor

3.5” floppy drive; CD-RW CD-ROM drive

Windows

®

2000 operating system

Microsoft

®

PS/2 IntelliMouse

DuoFlow software, version 5.0 or higher

Networkable for printer and file access

Maximizer

Size: 18.5” x 14.25” x 4.33”; 47 x 36 x 11 cm (L x W x H)

Weight: 11 lbs; 4.8 kg

Environmental: Cold room compatible

Solvent compatibility: 1M sodium hydroxide (NaOH), hydrochloric acid (HCl), and organic acids;
0.1M sulfuric acid; 7M urea and guanide hydrochloric acid; 100% methanol (MeOH), isopropanol
(IPA), and acetonitrile; 0.1% triflouracetic acid (TFA); 1% detergents (including sodium dodecyl
sulfate (SDS)) and TritonX-100

Workstation pump

Dual piston, positive displacement solvent pumps, with interchangeable solvent delivery pumpheads
(F10 and F40) made of bio-compatible PEEK

F10 pumpheads: Without Maximizer: 3,500 psi (233 bar, 23 MPa) from 0.01 to 10.0 ml/min in 0.01
ml/min increments. With Maximizer: 3,500 psi (233 bar, 23 MPa) from 0.5 to 20.0 ml/min in 0.02
ml/min increments.
Flow accuracy: 0.01 to 10.0 ml/min: 2%
Flow reproducibility: 0.01 rsd

F40 pumpheads: Without Maximizer: 1,000 psi (66 bar, 6.6 MPa) from 0.5 to 40.0 ml/min in 0.01
ml/min increments. With Maximizer: 1,000 psi (66 bar, 6.6 MPa) from 1.0 to 80.0 ml/min in 0.02
ml/min increments
Flow accuracy: 0.5 to 40.0 ml/min: 2%

Compositional accuracy: ±1.0%

High pressure dynamic mixing. MX-1 mixer volume without barrel: 263 µl; mixer without mixer barrel
extender: 750 µl; mixer with mixer barrel extender: 2.0 ml. Maximizer mixer volume without barrel:
750 µl; mixer with mixer barrel extender: 5.75 ml; mixer with expansion barrel: 12.75 ml.

Built-in pressure transducer

Gradient by software control: 1% increments

Advertising