GENERAL GUIDELINES FOR CREATING YOUR OWN TUBING CONNECTIONS

The DuoFlow system uses three types of tubing. Use the following table to select the appropriate tubing.

The Fittings Kit comes with fittings, tubing, and an F10 Tubing Kit, with the necessary tubing cut and
fitted for each system connection (see Figure 4-1). Complete description of each of these items is
provided with the Fittings kit.

Table 4-1.

Tubing Guidelines

Use

Tubing Dimensions

Tubing Material

Fittings

Vol/cm

Pre-pump*

1/8” OD x 0.062” ID

clear,

PTFE

Super flangeless fittings

20 µl

(Workstation

(3.2 mm OD x 1.6 mm ID)

for 1/8” (3.2 mm) OD tubing

and Aux pumps)

Post-pump**

1/16 OD x 0.02” ID

clear, Tefzel

Super flangeless fittings

2 µl

(1.6 mm OD x 0.5 mm ID)

for 1/16” (1.6 mm) OD tubing

Post-pump**

1/16” OD x 0.020” ID

orange, PEEK

1/4-28 Super flangeless fittings 2 µl

with F10

(1.6 mm OD x 0.5 mm ID)

for 1/16” (1.6 mm) OD tubing

pumphead

Post-pump**

1/16” OD x 0.030” ID

green, PEEK

1/4-28 Super flangeless fittings 4.5 µl

with F40

(1.6 mm OD x 0.76 mm ID)

for 1/16” (1.6 mm) OD tubing

pumphead

* Pre-pump refers to all tubing leading up to pump inlet port. Pumps include Workstation pumps, EP-1 pump, and Econo Gradient Pump. Maximizer uses

pre-pump tubing.

**Post-pump refers to all tubing following the pump outlet port.

When plumbing the system, be sure to keep tubing lengths to a minimum. All fittings should be finger-tight.
Do not over-tighten, as you risk damaging the connection.

Ferrule Installation: 1/8” (3.2 mm) OD and 1/16” (1.6 mm) OD tubing

1.

Cut the tubing with the tubing cutter provided in the Fittings Kit. This will give a flat, clean cut to the
tubing; this is important in making a fitting that does not leak.

Figure 4-2. Making 1/4-28 flat bottom fittings

TUBING

CUT THE

TUBING

NUT

STAINLESS STEEL

COMPRESSION RING

PEEK FERRULE

FITTINGS TIGHTENER

FITTINGS

TIGHTENER

SYSTEM PLUMBING

SYSTEM INSTALLATION AND SETUP

4-2

Advertising