Table 2-9.

Maximizer Screens

Function and Description on Maximizer Faceplate in Local Mode

Inlet Selection

Arrow buttons switch between Inlets A1(0) and A2(1).
PREV changes to previous screen.
NEXT changes to next screen.

Arrow buttons switch between Inlets B1(0) and B2(1).

Valve Control

Arrow buttons select a port <Port> on valve <Valve>. Up to six valves
may be displayed including three motorized (AVR7-3, AVR9-8) and three
solenoid (SVT3-2, SV5-4) valves.
ENTER accepts the change and moves the valve to the new position.

SIM Calibration

Displays the current Maximizer SIM voltage.

Used in conjunction with a 1 Volt calibration source.
Pressing ENTER sets the current voltage reading to 1 V.
Pressing the UP arrow resets the calibration to the factory setting.

After ENTER is pressed, the display responds with “SET” to show that the
calibration was successful.

After the UP arrow is pressed, the display responds with “RESET” to
show that the calibration was reset.

pH Calibration

Displays the current pH and temperature.

ENTER causes the Maximizer to enter pH calibration mode.

DESCRIPTION OF BIOLOGIC DUOFLOW SYSTEM

SYSTEM OVERVIEW

2-16

Screen

CAL pH?
PREV

ENTER = Y

NEXT

pH 6.00
PREV

22.6 C
NEXT

CAL SIM?
PREV reset

ENT=SET

NEXT

CAL SIM?
PREV set

ENT=SET

NEXT

CAL SIM?
PREV

ENT=SET

NEXT

SIM 0.000
PREV

Volts
NEXT

<Valve><Port>
PREV

POS 0
NEXT

BLEND B
PREV

POS 1
NEXT

BLEND A
PREV

POS 0
NEXT