7.2.2 Inlet and Valve Naming
One of the DuoFlow Software's most powerful features is the ability to define a valves function and to name
inlet and valve positions. This greatly simplifies programming individual steps in the Protocol Editor. Each
DuoFlow inlet (A, B, or A1, A2, B1, B2 on a system configured with a Maximizer) can be assigned to a
specific valve (Inlet Valve) or be assigned to a specific buffer. The "Gradient Pump" and "Maximizer +
Gradient Pump" section of the Setup screen is used to assign Inlets A and B (A1, A2, B1 and B2 on a
system with a Maximizer) to a specific buffer. Alternatively, the SVT3-2, SV5-4 and AVR9-8 valves can be
assigned as Inlet valves and each port of the inlet valves assigned to a specific buffer.