Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 216

Advertising
background image

Problem

Possible Cause

Solution

False or “ghost”

The injection valve, injection

peaks.

port and loop were not rinsed
out between sample injections.

Strongly retained sample
components may still be
eluting from a previous run.

12.4 TROUBLESHOOTING THE CONDUCTIVITY FLOW CELL AND TRACE

Problem

Possible Cause

Solution

No values for

There may be a loose or

buffer conductivity

incorrect cable connection.

are displayed on
the Controller
status bar.

Conductivity monitor may
need calibration.

TROUBLESHOOTING

MAINTENANCE AND TROUBLESHOOTING

12-9

1. It is a good idea to allow a larger volume of

injection buffer to pass through the inject valve
and loop. For example, if the sample loop
contains 100 µl, then program a load/injection
volume of 200 µl to completely flush the valve
and loop.

2. Always clean large volume dynamic loops, such

as the Bio-Rad DynaLoop, between use to
avoid cross contamination.

3. Use stringent elution conditions or consult the

column's user manual for an appropriate wash
step to remove such components between
runs.

1. Make sure the cable from the Conductivity flow

cell is plugged into the correct connector on the
rear of the Workstation and that the pins are
not bent.

2. Use the Conductivity Flow Cell Constant

Calibration function from the Utilities drop-
down menu. Either input the constant value
which is shown onthe tag attached to the flow
cell or calibrate the flow cell using 0.5M sodium
chloride (NaCI). Note that the relationship
between the conductivity reading and salt
concentration is not linear. The curve will flatten
out at higher salt concentrations.

3. If a Maximizer is connected, make sure the

Conductivity monitor is plugged into it.

Advertising