3.5

MIXERS

Two mixers have been designed for use with the DuoFlow system: the MX-1 mixer and the higher capacity
Maximizer mixer. The mixers are shipped pre-assembled with their mixer barrel. The mixer barrel may be
removed or replaced increasing or decreasing the internal volume to provide the mixer capacity appropriate
to the flow rate. Refer to table 3.1 when selecting a mixer volume.

Figure 3-7. Mixers

Table 3-1.

Mixer Flow Rates

Flow Rate

Barrel Extension

Capacity

Assembly Screws

MX-1 Mixer (for use without Maximizer)
less than 1ml/min

none

263 µl

10-32 x 5/8” (1.6 cm)

1 to 10 ml/min

750 µl mixer barrel extender

750 µl

10-32 x 7/8” (2.2 cm)

10 to 40 ml/min

2 ml mixer barrel extender

2 ml

10-32 x 1-1/2” (3.8 cm)

Maximizer Mixer (for use with Maximizer)
0.5 to 10 ml/min

none

750 µl

1/4-20 x 1/2” (1.3 cm)

10 to 40 ml/min

5 ml mixer barrel extender

5 ml

1/4-20 x 1-1/2” (3.8 cm)

40 to 80 ml/min

12 ml mixer barrel extender

12 ml

1/4-20 x 3-1/4” (8.2 cm)

SCREWS

MIXER

BODY

O-RING

750

µ

l MIXER

BARREL EXTENDER

OUTLET PORT

O-RING

INLET PORT

MIXER

TOP

MAGNETIC
STIR BAR

O-RING GROOVE

O-RING GROOVE

ASSEMBLY OF THE MX-1 MIXER

MIXER

TOP

SCREWS

MIXER

BODY

12 ml MIXER
BARREL

OUTLET PORT

5 ml MIXER
BARREL EXTENDER

INLET PORT

MAGNETIC
STIR BAR

O-RING

O-RING GROOVE

O-RING GROOVE

ASSEMBLY OF THE MAXIMIZER MIXER

SYSTEM SETUP

SYSTEM INSTALLATION AND SETUP

3-7