Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 192

Advertising
background image

REVERSE FLOW CHROMATOGRAPHY

ADVANCED SYSTEM APPLICATIONS

10-5

(Optional) Check the pH Correction
9. From the manual screen press the Buffer Blending "Setup" button and select the desired buffer

system from the Buffer System list. Turn pH correction on and enter the Desired and Observed pH
from step 8 above.

10. Set the pH, Salt composition and flow rate. Press run.

11. After the system has equilibrated collect some effluent and measure the pH.

Figure 10-1. Buffer Blending Setup Dialog for a Single Component Buffer

Figure 10-2. Buffer Blender Setup Dialog for a Multi-Component Buffer

Two point correction (best for gradient experiments)

1. From the manual screen Press the Buffer Blender "Setup" button and select the desired buffer

system. Make sure the pH correction box is not checked.

2. Set the pH and flow rate. Set the salt composition to 0 %B. Press run.

3. After the system has equilibrated collect some effluent.

4. Set the salt composition to the maximum %B that will be used in the experiment. Press run.

5. After the system has equilibrated collect some effluent in a second tube.

Formate/Acetate/Phosphate

pH 3.10 to 6.80

Buffer Blender

Buffer System

Buffer System: Formate/Acetate/Phosphate

Sort By:

Name

pH Range

pH Range (at 25 Celcius)

Molarity at 100%B:

Temperature Compensation Coefficient:

pH Correction Values

Use pH Correction

3.10 to 6.80

Desired pH pH at 0%B pH at

%B

Buffer

100

Acetate

4.40

4.62

4.76

Formate

3.29

1.60

3.75

Phosphate

6.71

7.04

7.20

1.0

M

0.0002

Edit Buffer Blender

*Tris (25 mM)

pH 7.1 to 9.1

Solution A1: Acetic Acid (40 mM)
Prepare by dissolving: 80 ml of 1M Acetic Acid in water and diluting to 2.0 L.

Solution A2: Sodium Acetate Teihydrate (40 mM)
Prepare by dissolving: 10.9 g of Sodium Acetate Trihydrate (MW 136.1 g/mol) in water and diluting to 2.0 L.

Solution A3: Sodium Chloride (2.0 mM)
Prepare by dissolving: 233.6 g of Sodium Chloride(MW 58.4 g/mol) in water and diluting to 2.0 L.

Cancel

Buffer Blender

Buffer Blender

Buffer System

Sort By:

Name

pH Range

pH Range (at 25 Celcius)

Molarity at 100%B:

Temperature Compensation Coefficient:

Recipe

OK

pH Correction Values

Use pH Correction

Pump Head Selection: F10

Flow Rate Range:

0.20 - 20.0 ml/min

Max Pressure:

700

7.10 to 9.10

Desired

pH

4.76

Observed at 0%B

pH

4.70

Observed at

0%B

100

pH

4.48

1.0

M

-0.0280

psi

Use two point correction

Advertising