Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 209

Advertising
background image

Problem

Possible Cause

Solution

Instrument face-

A communication problem

plates (eg. pump,

has occurred.

UV, fraction collec-
tor) “drop-off” the
screen when in
Manual mode.

Controller monitor

The monitor or the

displays unusual

Controller video card

colors or characters. may be defective and need

replacement.

No sound from

Cables not connected

monitor speakers.

properly.

“Mute” switch is activated.

Controller displays

A communication problem

“General Protection has occured.
Fault” message or
“locks up” during
normal operation.

BioFrac Fraction

The BioFrac is not in Local

collector will not go

mode and in the main

into system mode

screen.

A method or run is

The file is open in either

colored red in the

the BioLogic Online or

browser and I

Offline window.

cannot open it.

The BioLogic Hardware
server is hung.

TROUBLESHOOTING

MAINTENANCE AND TROUBLESHOOTING

12-2

1. Exit the DuoFlow application and turn off the

Controller. Turn off the Workstation.

2. Make sure all Instrument Bus (phone jack style)

connections are secure in all the instruments. Turn
the Controller and Workstation back on.

3. If the problem persists, contact Bio-Rad Technical

Service.

1. Exit the DuoFlow application and restart the

Controller.

2. If the problem persists, contact Bio-Rad Technical

Service.

1. Consult the monitor instruction manual to ensure

that the cables are correctly connected.

2. Press the “mute” button, which is located on the

front of the monitor to turn on the speakers.

1. Exit the DuoFlow application and restart the

Controller.

2. If the problem persists, contact Bio-Rad Technical

Service.

Placed the BioFrac in Local mode so that it displays
the main screen (start-up screen). Press the System
option in the BioFrac panel of the BioLogic software's
manual screen.

Go to the BioLogic Online or Offline window where it is
open and from the File menu select File/Close.

Exit the BioLogic software, press Alt-Ctrl-Del and then
select the "Task List..." button. Select the Processes
tab and kill any of the following processes that may be
running: BL_HWSRV.exe, biologic.exe, _biolog_offline.
Restart the BioLogic software.