f.

Using the four long mounting screws removed earlier, remount the pumphead assembly to the
Workstation so that it is secure. Insert and alternately tighten the mounting screws. Do not over
tighten.

g. Using the priming port on the front of the pumphead, purge the pump with 50% methanol and

then flush with water.

3.

Care of the Pistons.
To maintain pump performance, the pumphead should be rinsed on a nightly basis. The pump may
be rinsed automatically by programming a water rinse step at the end of each protocol. It is also
important to rinse behind the piston seals by injecting water through the washout port, located at the
top of the pumphead. A standard syringe may be used for this purpose.

Figure 11-3. Pumphead Assembly

"SHORT"

SCREWS

(8-32 X 1.0")

"LONG"

SCREWS

(8-32 X 1.5")

NOTE: FOR BOTH "SHORT" AND "LONG" SCREWS
USE PUMPHEAD DISASSEMBLY ALLEN WRENCH
(OR STANDARD 9/64" ALLEN WRENCH)

OUTLET BLOCK

WASHOUT PORT

O-RING

O-RING

CENTER BLOCK

FLOW RATE

IDENTIFICATION

("10" or "40")

INLET BLOCK

CHECK VALVES

CHECK VALVES

PUMP HEAD