SAMPLE LOADING

ADVANCED SYSTEM APPLICATIONS

9-2

Writing the Column Switching Protocol

When using multiple columns, care should be taken to prevent cross-contamination. Purge steps should be
included before placing a column inline to ensure that the tubing is clean and filled with the appropriate
buffer. The following three steps should be included as part of a column-switching step.

1. Add a Change Valve step to your protocol and place the Column Switcher valve in the Purge

position.

2. Add an isocratic flow step to wash the tubing and mixer with 3-5 volumes of buffer.

3. Add a Change Valve step to your protocol and select the desired column on the Column Switcher

valve.

Column-switching steps are usually placed at the beginning of a protocol or at the start of each dimension in
a multi-dimensional chromatography experiment.

9.1.2 AVR9-8 Eight Column Switching

Figure 9-2. Column Switching using Two AVR9-8 Valves

For this application, two AVR9-8 valves are required.

UV DETECTOR AND

FRACTION COLLECTOR

AVR9-8 AS COLUMN
SWITCHING VALVE
TO RECEIVE SAMPLE

AVR7-3 INJECT VALVE

FOR SAMPLE APPLICATION

COLUMN INLET

SECOND AVR9-8 AS
COLUMN SWITCHING VALVE
TO DIRECT ELUENT,
TO DETECTORS AND
FRACTION COLLECTOR

COLUMN OUTLET

5

4

3

2

1

7

6

9

UP TO 8

DIFFERENT

COLUMNS

8

5

4

3

2

1

7

6

9

8

COLUMN

WORKSTATION
PUMP

SAMPLE
LOOP

SAMPLE

INJECT

WASTE

WASTE

6

7

1

2

3

4

5

INJECT