Modes of operation system operation 7-21 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 150

Advertising
background image

Table 7-10. (continued)

Fraction Collection

Description: Threshold

Threshold: This method allows the flow to be sent to the collector if the detector signal of the eluant is
above or below a certain signal level. Note: You can select any detection source to initiate threshold-based
fraction collection. Non-peak eluant can be sent to waste or to the fraction collector.

Fraction Size, Start Tube, End Tube, Start Rack and End Rack: See description for Collect All on
page 7-20.

Threshold: Enter the threshold value. Fractions will be collected whenever the detector output goes
above or below the threshold level.

Above/below Threshold: Used to select whether fractions are collected when the detector signal is
above or below a designated Threshold.

Non-peak parameters Destination and Fraction Size: Enter the non-peak parameters. If the buffer
stream is diverted to Waste, there is no need to enter a fraction size. But if you want to collect non-peak
material, you must enter a fraction size.

Collect All

Threshold

Collection Windows

Threshold and Collection Windows

Fraction Collection Scheme

BioFrac Rack Type: F1 (12-13 mm tubes)

Close

Waste

Destination:

Non-peak Parameters:

Tubes

Below Threshold

Above Threshold

A

B

90

Rack:

1

1A

1F

Tube #:

Row Col.

Start:

End:

1.00

Fraction Size

ml

0.100

Threshold

1.00

Fraction Size

ml

UV Detector

Threshold Detector

AU

MODES OF OPERATION

SYSTEM OPERATION

7-21