MAINTENANCE

MAINTENANCE AND TROUBLESHOOTING

11-7

11.4.2 Replacing the Lamp in the UV Detector

The following information covers the replacement procedures both for a new mercury lamp unit and a new
zinc lamp unit. Note that the zinc lamp assembly has cooling fins and is larger than the mercury lamp
assembly.

Indications that a lamp replacement is necessary include an unstable baseline and a decreased response to
a standard concentration of a test chromophore. In this latter case, it is advisable to ensure that the flow cell
is clean and that the correct wavelength filter is chosen before replacing the lamp. Refer to the previous
section.

Note: As indicated during the following procedure, make sure the DuoFlow Workstation is turned off.
This will eliminate the risk of electric shock, thermal burns, or UV light exposure to you or damage to the
detector during cable connection.

Figure 11-4. Replacing a Mercury Lamp in a Model OM-II UV Detector

Figure 11-5. Replacing a Zinc Lamp in a Model OM-II UV Detector

ZINC LAMP
HOUSING

FILTER

FLOW CELL

ZINC LAMP

FIXTURE

MERCURY
LAMP HOUSING