Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 140

Advertising
background image

Table 7-3.

Isocratic Flow

The Edit Isocratic Flow dialog is used to define the isocratic step buffer composition. The buffer composition
is determined by the percentage of total buffer delivered through pumps A (%A) and B (%B).

Buffers: Displays all the available buffers. The names displayed for Buffer A and Buffer B are the
names assigned to the pump inlets or inlet valves in the device setup (see Chapter 7.2.2).

Composition: Allows you to set the buffer composition for the step (e.g., 75% A and 25% B).

The Edit Isocratic Flow pH View dialog is used to define the isocratic step buffer composition when a
Maximizer is connected to the system and Buffer Blending is turned on. The dialog box on the left lists the
current buffer system, its pH range and the solutions assigned to each Maximizer valve inlet.

pH: Allows you to enter the pH for the current step.

%B: Allows you to set the buffer composition for the step as a percentage of the 1x salt
concentration.

The following selections are available in both the Edit Isocratic Flow screen and the Edit Isocratic Flow
pH View
screen.

Volume/Time: Allows you to choose the duration of the step.

Flow rate (ml/min): Allows you to choose the flow rate of the step.

OK: Adds the step to the protocol. This is the same as pressing the Enter key on the keyboard.

Cancel: Cancels all input. This is the same as pressing the Esc key on the keyboard.

Step, Time or Volume: Identifies current step number, and calculates the elapsed time or volume from
all previous steps. This is not user editable.

Isocratic
Flow

Edit Isocratic Flow

Step 2, Volume = 0.00 ml

Isocratic Flow

Buffer A1

OK

Cancel

100

A

Buffer B1

Buffers

% Composition

0

1.0

Volume (ml.)

1.00

B

Flow (ml./min)

Edit Isocratic Flow pH View

Step 2, Volume = 1.00 ml

Isocratic Flow pH View

Buffer System: Tris (25 mM)
pH Range = 7.20 to 9.20

Solution A1: Tris.HCI (50 mM)
Solution A2: Tris (50 mM)
Solution B1: Degassed Water
Solution B2: Sodium Chloride (2.00M)

Molarity at 100%B = 1.0 M

OK

Cancel

1.0

Volume (ml.)

1.00

Flow (ml./min.)

7.00

pH

0

%B

Isocratic Flow Screen

Isocratic Flow Screen with Maximizer in Use

MODES OF OPERATION

SYSTEM OPERATION

7-11

Advertising