Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 221

Advertising
background image

SPECIFICATIONS

APPENDIX A

A-2

USB Bitbus Communicator

Size: 4.25” x 2.75” x 1.125”; 11 x 7 x 3 cm (L x W x H)

Weight: 6 oz

Electrical: 5 VDC output

General

System fittings: Flangeless 1/4-28 flat-bottom fittings, fingertight nuts and ferrules

System Tubing for flow rates up to 15 ml/min: 1/16” (1.6 mm) OD, 0.020” (0.51 mm) ID orange
PEEK tubing. The approximate volume of 1 cm of tubing is 2 µl.
System Tubing for flow rates of 15 ml/min and greater: 1/16” (1.6 mm) OD, 0.030” (0.76 mm) ID
green PEEK tubing. The approximate volume of 1 cm of tubing is 4.5 µl.
Pump Inlet Tubing: 1/8” (3.2 mm) OD, 0.062” (1.6 mm) ID

PTFE tubing

Wetted materials: 100% bio-compatible

Solvent compatibility: 1.0 M sodium hydroxide, hydrochloric acid and organic acids, 1.0 M sulfuric
acid, 7M urea and Guanidine hydrochloric acid, 100% ethanol, isopropanol, and acetonitrile, 0.1%
trifluoroacetic acid (TFA), 1% detergents (including sodium dodecyl sulfate (SDS), Triton X-100)

Sound level, Workstation: <70 dBa

Dimensions, Workstation (approx.): ~35 x 41 x 18 cm (W x D x H); height with 3-tray rack: 76 cm
Dimensions, Maximizer (approx.): 36 x 47 x 11 cm (W x D x H)

Weight (Controller, Workstation, monitor, rack, inject valve and detection modules): 34.5 kg (76 lbs)
(approximate). Weight, Maximizer: 4.8 kg.

Operating temperature (excluding Controller and USB Bitbus Communicator): 4° – 40°C

Detection

Fixed wavelength, mercury lamp with 280/254 nm filter, auto zero; 0.0001 - 2.0 AUFS detection
range

Analytical (5 mm pathlength, 16 µl total volume) flow cell

Preparative (2 mm pathlength 30 µl volume) flow cell

Optional 214 nm detection kit (includes zinc lamp, housing, and 214 nm filter)

Optional wavelength filters include: 214, 313, 365, 405, 436, and 546 nm

Conductivity detection (500 mS/cm; 6 µl flow cell; accuracy ± 2% full scale)

Optional flow-through pH monitor (includes flow cell, pH electrode, and SIM)

Cold room compatible

Sample Loading

Fixed sample loops, 100 µl to 5.0 ml; automated injection valve included

Large volume loading through the Gradient Pump. Optional inlet select valves (SV5-4, AVR9-8) for
sanitation and multiple sample use.

Variable volume sample loading through the DynaLoop sliding-piston sample loop.

Software control of an auxiliary pump for multiple sample loop fills and large volume dynamic loop
filling.

Valve Control

Up to three low-pressure, automated solenoid valves (SV5-4 and SVT3-2) for stream selection.

Up to three high-pressure, automated valves (AVR7-3 and AVR9-8) for sample inject, column
switching and stream selection.