System setup system installation and setup 3-20 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 92

Advertising
background image

3.11.3 BioLogic Configuration Utility Software

The BioLogic Configuration utility is used anytime the pumpheads are changed or a Maximizer is installed. It
is also used to choose between Bio-Rad’s BioFrac and Model 2128 fraction collectors. To run this software:

1.

Exit the BioLogic DuoFlow software by selecting Exit from the File drop-down menu.

2.

Double-click on the BioLogic Configuration icon.

3.

In the BioLogic Configuration Utility window, indicate the pump that will be used with the system,
whether or not the Maximizer will be used with the system, and the default Bio-Rad fraction collector
(either the BioFrac or the Model 2128).

4.

Exit the BioLogic Configuration Utility and double-click on the BioLogic DuoFlow icon to start the
software.

Figure 3-19a. BioLogic Configuration Utility Software Screen

BioLogic Configuration Utility

Cancel

Help

OK

Flow Rate Range:

Max Pressure:

0.02 - 20.00 ml / min

3500 psi

Workstation Configuration

Pump

Spec

Purge Flow Rate:

High Pressure Limit:

Low Pressure Limit:

10.00 ml / min

3500 psi

0 psi

Setting

Maximizer is installed

Pump Head

Model 2128

BioFrac

F10

F40

Fraction Collector

SYSTEM SETUP

SYSTEM INSTALLATION AND SETUP

3-20