System setup system installation and setup 3-16 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 88

Advertising
background image

3.9

PUMP CONNECTIONS

This section discusses the connections for the following instruments and devices:

Model EP-1 Econo pump

Econo Gradient Pump (EGP)

3.9.1

Model EP-1 Econo Pump

The EP-1 Econo pump is connected to the system as discussed below:

To connect the EP-1 Econo pump with the DuoFlow system, make the following cable connections:

Figure 3-14. Connecting an EP-1 Econo Pump to the BioLogic DuoFlow Workstation

1. On System Cable 7 locate the red and blue wires at the wire end of the cable and connect them to

the DuoFlow Workstation.
Cut short all other wires and insulate with tape.

a.

Connect the red wire to pin 6 of the Aux connector.

b.

Connect the blue wire to pin 9 of the Aux connector.

2. Plug the 8-pin mini-DIN connector into the Aux connector on the rear panel of the Econo pump.

PharMed tubing is recommended for use in the pumphead of the EP-1 Econo pump. For additional
information on this pump, consult its user manual. The illustration below shows the Aux pump connected to
port 3 of the Inject Valve.

ECONO PUMP

REAR PANEL

SYSTEM CABLE 7

RED TO PIN #6

BLUE TO PIN #9

WORKSTATION

REAR PANEL

AUX

UV
CHART

COND
CHART

UV
OPTICS

COND
FLOWCELL

SOLENOID VALVES

MIXER

1

2

3

UV LAMP

AUTOMATED VALVES

INSTR. BUS

4

5

6

1. INJECT
2. n/c
3. n/c
4. n/c
5. FC ADV
6. AUX PUMP
7. n/c
8. n/c
9. GND

10..25V 0.3A MAX.

SYSTEM SETUP

SYSTEM INSTALLATION AND SETUP

3-16