Table 7-8.

Hold

Inserts a Hold step into the method. The Hold step stops the progression of the method; the
pumps continue pumping at the Hold step %B composition. The method time (or volume
advances. When the run resumes, the current fraction collection condition is maintained. The
hold condition is maintined until a specified activity occurs (e.g., “Hold until Start of Inject” or
“Hold until Keypress”). For example, if the hold starts at time 2 (minutes), then the method
resumes at time 2, no matter how long the Hold was in effect.

Hold Until and Threshold Detector: Choose one of five events to discontinue the hold:

Key Pressed: Press the F2 key on the keyboard to end the programmed hold.

Start of Inject

:

The run will continue when the manually-controlled device (which must be connected to

the Workstation AUX connector at pin 1, Inject) is moved to its desired position.

End of Inject: The run will continue when the manually-controlled device (which must be connected to
the Workstation AUX connector at pin 1, Inject) is returned to its original position.

Above Threshold: The run will continue when the selected detector signal exceeds the specified
threshold:

Threshold Detector: Select the desired threshold detector absorbance, conductivity, RI, etc.

Threshold: Specify a threshold value.

Below Threshold: The run will continue when the selected detector signal falls below the specified
threshold:

Threshold Detector: Select the desired threshold detector absorbance, conductivity, RI, etc.

Threshold: Specify a threshold value.

Time Out Reqd and Time Out (min): To hold for a specified length of time.

Sound Alarm: When this box is checked, an alarm will sound at the beginning of the step to remind you
that the system requires an action on your part to allow the method to advance.

OK: Adds the step to the protocol. This is the same as pressing the Enter key on the keyboard.

Cancel: Does

not add the step to the protocol. This is the same as pressing the Esc key on the

keyboard.

Step, Time or Volume: Identifies current step number, and calculates the elapsed time or volume from
all previous steps. This is not user editable.

Hold

Edit Hold Until Event

Step 1, Volume = 0.00 ml

Hold Until Event

OK

Cancel

UV Detector

1.000

Time Out (min)

The method will automatically
continue after the selected
time-out duration

Threshold

Key Pressed

Time Out Reqd.

Sound Alarm

Start of Inject

End of Inject

Above Threshold

Below Threshold

Hold Until

Threshold Detector

AU

MODES OF OPERATION

SYSTEM OPERATION

7-18