Table 2-9. (continued)

Maximizer Screens

Function and Description

pH Calibration (continued)

CURSOR moves the cursor to the next digit of the pH set point.
UP/DOWN arrows adjust the pH set point.
CANCEL aborts the calibration.
ENTER accepts the pH value of the first calibration buffer and moves to
the next screen.

Display of the current pH voltage for monitoring probe equilibration.
SET or ENTER sets the first calibration point.
CANCEL aborts the calibration.

CURSOR moves the cursor to the next digit of the pH set point.
UP/DOWN arrows adjust the pH set point.
CANCEL aborts the calibration.
ENTER accepts the pH value of the second calibration buffer and moves
to the next screen.

Displays the current pH voltage for monitoring probe equilibration.
SET or ENTER sets the second calibration point and exits pH calibration
mode.
CANCEL aborts the calibration.

Conductivity Calibration

Displays the current conductivity reading (mS/cm).

ENTER causes the Maximizer to enter calibration mode.

CURSOR moves the cursor to the next digit of the calibration constant.
ENTER accepts the conductivity cell constant, calibrates the conductivity
meter, and exits calibration mode.
CANCEL aborts the calibration.

Beeper

ENTER causes the Maximizer to enter volume adjustment mode.

The UP and DOWN arrows adjust the volume of the alarm.

Firmware Version

Displays the current version of the Maximizer firmware.

DESCRIPTION OF BIOLOGIC DUOFLOW SYSTEM

SYSTEM OVERVIEW

2-17

Screen

MAXIMIZER

V3.xx

NEXT

VOLUME: UP/DOWN

ENTER TO FINISH

BEEPER?
PREV

ENTER=Y

NEXT

NEW CELL C:
CURSOR

43.00

CANCEL

SET CELL C?
PREV

ENT=Y

NEXT

COND.
PREV

0.86
NEXT

2ND:
CANCEL

35.22 mV

SET

2ND pH PT.
CURSOR

_7.00

CANCEL

1ST:
CANCEL

35.22 mV

SET

1ST pH PT.
CURSOR

_4.00

CANCEL