Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 49

Advertising
background image

The BioFrac collects fractions in a serpentine pattern for all racks but may be changed to a row or
column pattern for microplates and microtiter tubes. Fourteen collection choices are possible with a
total of nine racks. The BioFrac accommodates 12-20 mm and 30 mm tube diameters, microtubes,
and scintillation vials in a variety of diameters. The optional ice bath/microplate rack allows collection
into 13 mm diameter tubes, microplates, and microtiter tubes that meet SBS standards required for
automated microplate systems. The drop head height can be adjusted to accomodate tubes up to
150 mm.

The following racks are available.

Table 2-12.

BioFrac Racks Available

Rack Name

Description

Capacity (Format)

F1

12-13 mm diameter tube;

2 racks (6x15), 180 tubes

150 mm height

F2

15-16 mm diameter tube;

2 racks (5x12), 120 tubes

150 mm height

F3

18-20 mm diameter tube;

2 racks (4x10), 80 tubes

150 mm height

H1

1.5-2.0 ml capless microtubes

4 racks (6x7), 168 tubes

H2

0.5 ml capless microtubes

4 racks (7x9), 120 tubes

H3

16 mm scintillation vials

4 racks (5x6), 120 tubes

H4

30 mm scintillation vials

4 racks (2x3), 24 vials

H4-High

30 mm tubes

4 racks (2x3), 24 tubes

Ice Bath,

Ice bath for 13 mm diameter

1 rack (10x12), 120 tubes

Microplate rack

tubes, standard microplate or

4 plates (12x8), 96 wells

microtiter tubes (SBS standard

4 plates (8x6), 48 wells

format)

4 plates (4x6), 24 wells
4 plates (3x4) 12 wells

Prep-20

Preparative rack

bottle size fractions (20 funnels)

The BioFrac communicates with the DuoFlow Controller via the Instrument Bus. To connect the BioFrac to
the bitbus use either of the instrument bus connectors on the rear of the fraction collector. The BioFrac
fraction collector supports collection schemes such as Collect All, Threshold, Collection Windows, and
Threshold + Collection Windows (all with Delay volume, if required). The BioFrac diverter valve can be
mounted on either side of the fraction collector’s upright supports. See Section 3.8.1 for more information on
connecting the BioFrac to the DuoFlow system.

The BioFrac is plumbed to the DuoFlow system using 0.020” diameter PEEK tubing and 1/4-28 fittings for
flow rates up to 20 ml/min or with 0.030” diameter PEEK tubing and 1/4-28 fittings for flow rates up to
100 ml/min.

This instrument is described in detail in its separate documentation.

DESCRIPTION OF SYSTEM COMPONENTS

SYSTEM OVERVIEW

2-36

Advertising