Modes of operation system operation 7-20 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual

Page 149

Advertising
background image

Table 7-10.

Fraction Collection

Button

Description: Collect All

To set up fraction collection. Fraction collection schemes are discussed below.

IMPORTANT NOTE:

The DuoFlow system is designed to control the BioFrac and the Model 2128 fraction collectors, which are
the only collectors that provide a choice of racks and the ability to overlay fractions from consecutive runs.

The DuoFlow system will also accommodate the Bio-Rad Model 2110 fraction collector as well as generic
collectors. To use a fraction collection scheme other than “Collect All” with these collectors, it is

essential to

assign a SVT3-2 valve as a “Fraction Collector Diverter” in the Setup screen. Rack choice and overlay
features are

not

available with these collectors.

Collect All: The eluant from the entire run will be collected by the fraction collector.

Fraction Size: Enter the fraction size as volume (ml).

Tubes Required and Number of Racks: The displayed value is calculated based on the total volume
divided by fraction size.

Start Tube and End Tube: Identifies the first and last tubes to receive fractions. This applies only to the
BioFrac and the Model 2128 fraction collector.

Start Rack and End Rack: Identifies the first and last BioFrac rack to receive fractions.

Collect All

Threshold

Collection Windows

Threshold and Collection Windows

Fraction Collection Scheme

BioFrac Rack Type: F1 (12-13 mm tubes)

Close

Waste

Destination:

Non-peak Parameters:

Tubes

Below Threshold

Above Threshold

A

B

90

Rack:

1

1A

1F

Tube #:

Row Col.

Start:

End:

1.00

Fraction Size

ml

0.100

Threshold

1.00

Fraction Size

ml

UV Detector

Threshold Detector

AU

MODES OF OPERATION

SYSTEM OPERATION

7-20

Fraction

Collection