3.8

FRACTION COLLECTOR CONNECTIONS

This section discusses the connections for the following instruments and devices:

BioFrac fraction collector

Model 2110 fraction collector

Model 2128 fraction collector

Refer to section 2.7 for more detailed information about the fraction collectors.

3.8.1

BioFrac Fraction Collector

The BioFrac fraction collector is controlled by the DuoFlow software version 4.0 or greater via the
Instrument Bus. The BioFrac fraction collector is connected to the system as discussed below:

1.

Connect the USB Bitbus to the Controller as discussed in Section 3.2.

2.

Use either of the instrument bus connectors on the rear of the fraction collector to connect the
BioFrac fraction collector to the Instrument bus. (see Section 3.3)

3.

Select BioFrac in the BioLogic Configuration Utility.

3.8.2

Model 2110 Fraction Collector

The Model 2110 fraction collector may be connected to the Workstation, as discussed below:

1.

Connect the DB-9 connector on System Cable 5 to the Model 2110 fraction collector.

2.

Connect the bare wires to the AUX connector on the Workstation as follows: Black wire to pin 5 and
white wire to pin 9.

3.8.3

Model 2128 Fraction Collector

The Model 2128 fraction collector is controlled by the DuoFlow Controller via the Instrument Bus. The
Model 2128 fraction collector is connected to the system as discussed below:

1.

Connect the USB Bitbus to the Controller as discussed in section 3.2.

2.

Use the instrument bus connector on the rear of the fraction collector to connect the Model 2128
fraction collector to the instrument bus. Since the Model 2128 has only one instrument bus
connector, it should be the last device "daisy-chained" to the instrument bus.

3.

Select Model 2128 in the BioLogic Configuration Utility.

SYSTEM SETUP

SYSTEM INSTALLATION AND SETUP

3-15