Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual

Page 106

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-34

Table 4-7C. Ext Trig # > Calibration > Submenu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


AutoCal

Pulse, Mod & Stat

Initiates the automatic trigger channel

(Action)

START

calibration routine

Disconnect all signal sources from the
External Trigger input to be calibrated. Press
the Autocal START menu key to initiate a
zero and amplitude calibration.

Trigger channel zero and calibration
information is saved in a system file and
restored during initialization.

CALibration[3|4]:AUTO



Zeroing

Pulse, Mod & Stat

Initiates the automatic trigger channel

(Action)

START

zero adjustment


Disconnect all signal sources from the
External Trigger input to be zeroed. Press
the Zeroing START menu key to initiate
a zero.

CALibration[1|2]:ZERO








Figure 4-22

Ext Trig # > Calib