Sense:filter:time, Sense:filter:time -70, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 270: Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-70

SENSe:FILTer:TIMe

Description:

Set or return the current length of the integration filter on the selected channel. If the filter state is
set to AUTO, querying the time will return -0.01, and if set to OFF, a time query will return 0.00.
Note that setting the filter time will force the state to ON.

Syntax:

SENSe[1|2]:FILTer:TIMe <n>

Argument:

<n> = 0.002 to 16.000 seconds in 2 millisecond increments

Valid Modes:

Modulated

Advertising