Boonton 4500B RF Peak Power Analyzer

Operation

4-65

Table 4-16. Meas > Param Meas > Submenu (pulse)

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Selection

Pulse

Select channel for status or parameter display

(Mult. Choice)


Param Mode

Pulse

Select type of display

(Toggle)

Status, Meas

Select Status to display the following nine

parameters for the active channel in the field
above the grid box:
Freq, VidBW, Avging, TrA, TrB, Tr Dly,
Vscale, Vcent and Offset.

Select Meas to display nine selected
parameters in the display field above the
Grid box. The parameters are selected from
a list that includes the fifteen auto-measure
parameters as well as the status parameters.
The channel selection for each parameter is
determined by the active channel when the
selection is made. See Table 4-16A.

DISPlay:PARAmeter:MODE {MEAS,

STAT}




Param Column

Pulse

Select column in parameter display

(Mult. Choice)

L, M, R



Param Top

Pulse

Select top item in column

(Mult. Choice)

See Table 4-16A.



Param Middle

Pulse

Select middle item in column

(Mult. Choice)

See Table 4-16A.

Fig. 4-33.

Meas > Param Meas
menu


Param bottom

Pulse

Select bottom item in column

(Mult. Choice)

See Table 4-16A.

DISPlay:PARAmeter:TOP:LEFT {see Table 5-1}

DISPlay:PARAmeter:MIDdle:LEFT {see Table 5-1}

DISPlay:PARAmeter:BOTtom:LEFT {see Table 5-1}
Middle and Right columns similar to above. See Chapter 5.