Step 3: The Calibrator Mismatch Uncertainty is calculated using the formula in the previous section, using the 1 GHz
calibrator‘s published figure for ρ

CAL

and calculating the value ρ

SNSR

from the SWR specification on the 56518‘s datasheet.

ρ

CAL

= 0.091 (internal 1 GHz calibrator‘s reflection coefficient)

ρ

SNSR

= (1.15 - 1) / (1.15 + 1) = 0.070 (calculate reflection coefficient of 57518, max SWR = 1.15 at 1 GHz)

U

CalMismatch

= ±2 × ρ

CAL

× ρ

SNSR

Ч 100 %

= ±2 Ч 0.091 Ч 0.070 Ч 100 %

= ±1.27%

Step 4: The Source Mismatch Uncertainty is calculated using the formula in the previous section, using the DUT‘s
specification for ρ

SRCE

and calculating the value ρ

SNSR

from the SWR specification on the 56518‘s datasheet.

ρ

SRCE

= 0.057 (source reflection coefficient at 900 MHz)

ρ

SNSR

= (1.15 - 1) / (1.15 + 1) = 0.070 (calculate reflection coefficient of 56518, max SWR = 1.15 at 0.9 GHz)

U

SourceMismatch

= ±2 × ρ

SRCE

× ρ

SNSR

Ч 100 %

= ±2 Ч 0.057 Ч 0.070 Ч 100 %

= ±0.80%

Step 5: The uncertainty caused by Sensor Shaping Error for a 56518 peak sensor is 4% at all levels, from the sensor‘s
datasheet. But since we‘re measuring at 900MHz, which is very close to the 1GHz AutoCal frequency, we‘ll assume that the
frequency-dependent portion of the shaping error becomes very small, and we‘ll estimate that 2% remains.

U

ShapingError

= ±2.0 %

Step 6: The Sensor Temperature Drift Error depends on how far the temperature has drifted from the sensor calibration
temperature, and the temperature coefficient of the sensor. In our case, we are using a temperature compensated sensor, and
the temperature has drifted by 11 degrees C (49C - 38C) from the AutoCal temperature. We will use the equation in the
previous section to calculate sensor temperature drift uncertainty.

U

SnsrTempDrift

= ± (0.93% + 0.069% /degreeC)

= ± (0.93 + (0.069  11.0)) %

= ± 1.69%

Step 7: This is a relatively high-level measurement, so the noise contribution of the sensor is probably negligible, but we‘ll
calculate it anyway. We‘ll assume modulate mode with default filtering. The signal level is 13dBm, or 20mW. The ―noise
and drift‖ specification for the 57518 sensor is 50nW, from the sensor‘s datasheet. Noise uncertainty is the ratio of these two
figures.

U

Noise&Drift

= ± Sensor Noise (in watts) / Signal Power (in watts)

= ±50.0e-9 / 20.0e-3  100 %

= ±0.0003%

Step 8: A separate Sensor Zero Drift calculation does not need to be performed for peak sensors, since ―noise and drift‖ are
combined into one specification, so we‘ll just skip this step.