Marker:math:each, Marker:position:percent, Marker:math:each -46 – Boonton 4500B Peak Power Meter User Manual

Page 246: Marker:position:percent -46, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-46

MARKer:MATH:EACH

Description:

Set the marker group to EACH and configure the marker math function.

Syntax:

MARKer:MATH:EACH <arg1>, <arg2>, <arg3>

Argument:

<arg1> { RATIO, DELTA }

Specifies the result displayed in the center marker window

RATIO -the ratio of the left and right marker window readings in dBr or percent.

DELTA -valid only with <arg3> set to W (watts) or V (volts). The difference of the left
and right marker window readings expressed in channel units

<arg2> { MK1MK2, MK2MK1}

Specifies how the left and right marker windows are used to calculate the center marker
window.

MK1MK2 – ratio is MK1/MK2; difference is MK1-MK2

MK2MK1 - ratio is MK2/MK1; difference is MK2-MK1

<arg3> Specifies the channel units.

{dBm, W, V, dBV, dBmV, dBuV}

Valid Modes:

Modulated, Pulse and Statistical (RATIO with MK1MK2 only) modes

NOTE

There is no query form for MARKer:MATH:BOTH or MARKer:MATH;EACH.

MARKer:POSItion:PERcent

Description:

Set or return the percent probability (X-axis-position) of the selected vertical marker. Note that
the power value returned for each marker will depend on the setting of CALCulate:MODE.
When set to CDF, the highest power levels are towards the right side of the screen, with maximum
(highest peak) power occurring at 100%. When set to CCDF (also called 1-CDF), the highest
levels are towards the left, with peak power at 0%.

Syntax:

MARKer[1|2]:POSItion:PERcent <n>

Argument:

<n> = 0.000 to 100.000 %

Valid Modes:

Statistical modes only.