Calculate:math:arga, Calculate:math:argb, Calculate:math:operator – Boonton 4500B Peak Power Meter User Manual

Page 212: Calculate:mode, Calculate:math:arga -12, Calculate:math:argb -12, Calculate:math:operator -12, Calculate:mode -12, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-12

CALCulate:MATH:ARGA

Description:

Set or return the signal source of the A argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result

Syntax:

CALCulate:MATH:ARGA <asc>

Argument:

<asc> = {CH1 CH2, TRIG1, TRIG2, MEM1, MEM2}

Valid Modes:

Any

CALCulate:MATH:ARGB

Description:

Set or return the signal source of the B argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result

Syntax:

CALCulate:MATH:ARGB <asc>

Argument:

<asc> = {CH1 CH2, TRIG1, TRIG2, MEM1, MEM2}

Valid Modes:

Any

CALCulate:MATH:OPERator

Description:

Set or return the pixel-by-pixel mathematical operation performed between sources A and B to
create the math channel result. Note that the apparent operation performed is different for
logarithmic and non-logarithmic units for equivalent results.

Syntax:

CALCulate:MATH:OPERator <asc>

Argument:

<asc> = {CH_SUM, CH_DIFF, CH_RAT}

Log units

Non-log units

CH_SUM

------

add

CH_DIFF ------

subtract

CH_RAT subtract

divide

Valid Modes:

Any

CALCulate:MODE

Description:

Set or return the system measurement mode. MODULATED is a continuous measurement mode,
PULSE is a triggered, oscilloscope-like mode, and STATISTICAL, which includes CDF, CCDF
and PDF as sub-modes, may be operated in a start-stop continuous mode, a decimated continuous
mode or a gated (triggered) mode.

Syntax:

CALCulate:MODE <asc>

Argument:

<asc> = {MODULATED, PULSE, STATISTICAL}

Note: The STATISTICAL sub-mode CDF, CCDF or PDF is restored but not reported.

Valid Modes:

Any

Advertising