Boonton 4500B RF Peak Power Analyzer

Application Notes

6-14

Table 6-3. Statistical Automatic Measurements (Opt 10 only)

In the Statistical mode the following five automatic measurements are displayed in the Model 4500B Text display for both
input channels and both trigger channels. The Text column lists the abbreviated forms that appear on the display screen.

Text

TERM

DEFINITION


Avg

Average Power:

The unweighted avg of all power samples occurring since acquisition started.


Peak

Peak Power

The highest power sample occurring since acquisition was started.


Min

Minimum Power

The lowest power sample occurring since acquisition was started. In logarithmic
units a reading below the clip level will display as down arrows.

Pk/Avg

Pk/Avg Ratio

The ratio (in dB) of the Peak Power to the Average Power.


Dyn Rng

Dynamic Range

The ratio (in dB) of the Peak Power to the Minimum Power. Displays down

arrows if the minimum power is less than the clip level in log units.


The following six cursor measurements display the set position and measured value where the movable cursor intersects the
measurement trace. Note that the markers are undefined in PDF presentation format. The position and value text for each
marker or reference line will be displayed with the color of the assigned channel for that particular cursor.

Text

TERM

DEFINITION


RefLine1

Reference Line 1

Measures the CDF (or 1-CDF) % at the power level where reference line 1
intersect the assigned trace.

Note that if the intersection falls outside the horizontal screen extents, the
measured value will be clipped to % values within the horizontal axis. For this
reason, avoid zooming in horizontally with Time > X-Axis if you wish to make
Reference line measurements.

RefDelta

Reference Line Delta

For logarithmic units, displays the ratio (in dB) of the Reference Line 1 and
Reference Line 2 levels, or for linear units displays the difference (in %)
between the Reference Line 1 and Reference Line 2 levels and the difference (in
%) between their CDF (0r 1-CDF) values.


RefLine2

Reference Line 2

See Ref Line 1, above


Marker1

Marker 1

Measures the level at the selected %CDF (or 1-CDF).

Note that marker positions are rounded to the nearest pixel positions (1/50th of a
division), so for precise power measurements at very high or low % positions, it
may be necessary to zoom in horizontally with Statistics (Time) > Horz Scale.

Delta%

Marker Delta

Displays the difference (in %) between the marker positions and the ratio (in dB

for logarithmic units, % for linear units) of their levels.


Marker2

Marker 2

See Marker 1, above.