Table 4-9b. trig > menu (modulated), Table 4-9b. trig > menu (modulated) -49, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 121: Operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-49

Table 4-9B. Trig > Menu (modulated)

Menu Item

(Type)

Selections

Function


Trig Mode

Mod

Select the trigger system mode of operation

(Toggle)

Triggered, Free Run

Press the Trig > Trig Mode menu key to
select the desired mode of trigger system
operation.

Triggered – A trigger can be used to Start

the measurement, equivalent to pressing the
Run menu key. The measurement will
Continue until stopped with the menu key
or a bus command.

Free Run – Measurements are made

continuously without the need for a

Trigger.

TRIGger:MODE {TRIGGERED, FREERUN}


Trig Source

Mod

Same as Trig > Trig Source, Table 4-9 above

(Mult. Choice)

CH1, CH2,

Ext Trig1, Ext Trig2



Trig Level

Mod

Same as Trig > Trig Level , Table 4-9 above

(Numeric)

See Table 4-9



Trig Slope

Mod

Same as Trig > Trig Slope, Table 4-9 above

(Toggle)

+, -

Fig. 4-27e. Trig > menu

Modulated Mode