The following two global status values are displayed:

Text

TERM

DEFINITION

Total

Time

The total time in Hours:Minutes:Seconds that the data acquisition has been

running.

Samples

The total number of data samples that has been acquired for each channel in the
current run.

Note

The sampling speed of Channel 1 and Channel 2 is 25 MSa/sec. The sampling speed of the external

Trigger 1

and Trigger 2 inputs is 2 MSa/sec. If Channel 1 and Channel 2 are running, the sampling speed of each is

still 25

MSa/sec. If Channel 1 and Trigger 1 are running, the sampling speed for both is 2 MSa/sec. The same is

true for

Channel 2 and Trigger 2. For fastest sampling speed, do not run the CDF (or 1-CDF) of a channel and its

external

trigger at the same time.