Table 4-21A. Util > Remote > LAN Config > Submenu (Opt 11 only)


Menu Item

(Type)

Selections

Function


IP Address

None, info only

Displays the current LAN IP address. This
is configured by Windows network settings.

A display of 0.0.0.0 usually means that
either Util > Remote > Interface is set to
GPIB, or else it is set to LAN and the 4500B
is expecting an automatically assigned IP
address, but has not received one from a
DHCP server.


Network Conn

EXECUTE

Pressing this menu key will exit the 4500B

(Action)

application, and bring up the Windows™
―Network settings‖ dialog. Consult
Microsoft® documentation for details on
configuring network settings.

Navigating the dialog may be easier with
a keyboard and mouse connected, although
the 4500B‘s keyboard provides enough
functionality to perform certain configuration
tasks.

Once configuration is complete, close the
dialog and press the Init key to restart the
4500B application. If a new IP address has
been set or assigned, it should now be shown
in the IP Address menu box.

Fig. 4-37A. Util >

Remote Conf > LAN

Config > submenu