Fetch:marker:maximum, Fetch:marker:minimum, Fetch:marker:ratio – Boonton 4500B Peak Power Meter User Manual

Page 239: Fetch:marker:rdelta, Fetch:marker:maximum? -39, Fetch:marker:minimum? -39, Fetch:marker:ratio? -39, Fetch:marker:rdelta? -39, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-39


FETCh:MARKer:MAXimum?

Description:

For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.

Syntax: FETCh[1|2|3|4|6|7]:MARKer[1|2]:MAXimum?

Returns:

CC, maximum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

FETCh:MARKer:MINimum?

Description:

For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.

Syntax:

FETCh[1|2|3|4|6|7]:MARKer[1|2]:MINimum?

Returns:

CC, minimum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.




FETCh:MARKer:RATio?

Description:

For the specified channel return the ratio of MK1 to MK2. The units will be dB for logarithmic
units or percent for linear units..

Syntax:

FETCh[1|2|3|4|5|6|7]:Marker:RATio?

Returns:

CC, MK1/MK2

Where CC is the measurement condition code.

Valid Modes:

All

FETCh:MARKer:RDELTa?

Description:

For the specified channel return the difference between MK2 and MK1. The units will be the same
as marker units.

Syntax:

FETCh[1|2|3|4|5|6|7]:MARKer:RDELTa?

Returns:

CC, (MK2-MK1)

Where CC is the measurement condition code.

Valid Modes:

All

Advertising