Boonton 4500b rf peak power analyzer, Maintenance – Boonton 4500B Peak Power Meter User Manual

Page 348

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Maintenance

7-16




Table 7-3a. Calibrator Linearity - High Power Range

Cal Level Minimum Measured Maximum
(dBm) (dBm) (dBm) (dBm)

20.0 19.84 __________ 20.16

15.0 14.84 __________ 15.16

10.0 9.84 __________ 10.16



Table 7-3b. Calibrator Linearity - Low Power Range

Cal Level Minimum Measured Maximum
(dBm) (dBm) (dBm) (dBm)

5.0 4.9 __________ 5.1

-5.0 -5.1 __________ -4.9

-10.0 -10.16 __________ -9.84

-15.0 -15.16 __________ -14.84

-20.0 -20.25 __________ -19.75

-25.0 -25.25 __________ -24.75

-30.0 -30.25 __________ -29.75