Boonton 4500b rf peak power analyzer, Remote operation – Boonton 4500B Peak Power Meter User Manual

Page 303

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-103

Table 5-1. SCPI COMMAND SUMMARY (continued)


MEASure[1|2]:POWer?

Return modulated mode Pavg in dBm


MEASure[1|2]:VOLTage?

Return modulated mode Pavg in equivalent volts


MEMory:DRIVe Set/return active drive letter, A-Z

MEMory:FDOFfset:TBLA:DATA Set/return data array for frequency-dependent offset table A. N max = 60 pairs.

<asc> = Freq1, Ofset1, Freq2, Ofset2 ... FreqN, OfsetN

.

MEMory:FDOFfset:TBLB:DATA Set/return data array for frequency-dependent offset table B. N max = 60 pairs.

<asc> = Freq1, Ofset1, Freq2, Ofset2 ... FreqN, OfsetN


MEMory:FREE? Return the bytes of available memory free on selected drive

MEMory:OVERwrite Set/return the file overwrite state.

MEMory:SNSR[1|2]:CFFAST? Return sensor data

MEMory:SNSR[1|2]:CFSLOW? Return sensor data

MEMory:SNSR[1|2]:INFO? Return sensor data

MEMory:SNSR[1|2]:MESSage? Return sensor data

MEMory:SNSR[1|2]:TEMPCOMP?

Return sensor temperature compensation data


MEMory:SYS:LOAD Recall setup file and configure instrument. <arg> = filename with path but no

extension


MEMory:SYS:STORe

Build an instrument setup file and save to file. <arg> = filename with path but no

extension


MMEMory:COPY Copy from source argument to destination argument. (Ex: <args> = CH1 MEM2)

MMEMory:LOAD Read a reference waveform from disk into a reference channel. (Ex: <args> =

filename MEM2)


MMEMory:SAVE:DISPlay

Copy from display to destination argument.


MMEMory:STORe Write a waveform to disk from a specified channel. (Ex: <args> = TRIG1 filename)

OUTPut:LEVel:POWer

Set calibrator output level. <arg> = -50.0 to 20.0 dBm


OUTPut:LEVel:LIMit Set maximum power level for calibrator output. <arg> = -50.0 to 20.0 dBm

OUTPut:LEVel:STEP Set power step size for calibrator output level. <arg> = 0.1 to 60.0 dB

OUTPut:MODUlation Set calibrator to selected output mode. <arg> = CW, PULSE

OUTPut:POLArity Set calibrator edge to selected edge. <arg> = POS, NEG

Advertising