Remote Operation
5-20
DISPlay:HEADer
Description:
Set or return the type of header shown at the top left of the display.
Syntax:
DISPlay:HEADer <asc>
Argument:
<asc> = {BLANK (none), DATE (time and date), LOGO (instrument model), TEMP (sensors‘
internal temperature)}
Valid Modes:
All
DISPlay:HELP
Description:
Set the display to the show the Help page instead of Graphics or Text page. Help provides a short
description of each menu item. Set ON to see the HELP page; set OFF to restore the previous
GRAPH or TEXT page.
Syntax:
DISPlay:HELP <asc>
Argument:
<asc> = {ON, OFF}
Valid Modes:
All
DISPlay:MODE
Description:
Set or return the Text or Graphics page display setting. The Graphics page displays traces and
Marker measurements. The Text page displays auto-measure values in text form.
Syntax:
DISPlay:MODE <asc>
Argument:
<asc> = {GRAPH, TEXT}
Valid Modes:
All
DISPlay:MODUlated:TIMEBASE
Description:
Set or return the modulated mode timebase in seconds/division. The 4500B has fixed timebase
settings in a 1-2-5 sequence, and if the argument does not match one of these settings, it will be
forced to the next highest entry.
Syntax:
DISPlay:MODUlated:TIMEBASE <n>
Argument:
<n> = 5e-9 to 10 sec/div in a 1-2-5 sequence,
30 sec/div,
1 to 10 min/div,
30 min/div and 60 min/div
Valid Modes:
Modulated mode
Note:
There are separate time bases for the Modulated mode and the Pulse/Gated Statistical
modes. The arguments selected are saved and restored independently by mode.