16 system subsystem, System:autoset, System:beep – Boonton 4500B Peak Power Meter User Manual

Page 280: System:communicate:gpib:eoi, 16 system subsystem -80, System:autoset -80, System:beep -80, System:communicate:gpib:eoi -80, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-80

5.5.16 SYSTem Subsystem

The SYSTem group of commands is used to control system-level functions not directly related to instrument measurement
performance. SYSTem commands are used to return error codes or messages from the power analyzer error queue, control
hardware features (backlight and key beep), set the command language, and configure communication parameters for the
GPIB and serial interfaces.

SYSTem:AUTOSET

Description:

Perform the Auto Setup function to acquire a signal and display the trace in the Pulse mode.

Syntax:

SYSTem:AUTOSET

Argument:

None

Valid Modes:

Pulse mode is forced if not already set.

SYSTem:BEEP

Description:

Set or return the status of the audible keyboard beeper.

Syntax:

SYSTem:BEEP <asc>

Argument:

<asc> = {ON, OFF}

Valid Modes:

Pulse mode is forced if not already set.

SYSTem:COMMunicate:GPIB:EOI

Description:

Set or return the 4500B as TALKER EOI control parameter.

Syntax:

SYSTem:COMMunicate:GPIB:EOI <asc>

Argument:

<asc> = {ON, OFF}

Valid Modes:

Any