Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-54

5.5.11 OUTPut Subsystem

The OUTPut group of commands is used to control the internal 1GHz RF calibration source. The internal calibrator can be
configured as a programmable pulse source, with options for modulation rate and duty cycle. It can also be driven by an
external pulse modulation source.

OUTPut:LEVel:POWer

Description:

Set or return the power level for the 1 GHz calibrator output signal.

Syntax:

OUTPut:LEVel:POWer <asc>

Argument:

<asc> = -50.0 to +20.0 dBm (0.1dB resolution)

Valid Modes:

Any

OUTPut:LEVel:LIMit

Description:

Set or return the limit power level in dBm for the 1 GHz calibrator output signal. The calibrator
output level cannot be set to exceed this limit.

Syntax:

OUTPut:LEVel:LIMit <n>

Argument:

<asc> = -40.0 to +20.0 (0.1dB resolution)

Valid Modes:

Any

OUTPut:LEVel:STEP

Description:

Set or return the power step size in dB used by the Knob and Arrow keys on the front panel to
control the output level of the 1 GHz calibrator.

Syntax:

OUTPut:LEVel:STEP <n>

Argument:

<asc> = 0.1 to 60.0 (0.1dB resolution)

Valid Modes:

Any