5-56

OUTPut:PULse:PERiod

Description:

Set or return the pulse period in the Preset pulse control mode of the internal 1 GHz calibrator.
This parameter has no effect in the Variable pulse control mode External pulse source mode.

Syntax:

OUTPut:PULse:PERiod <arg>

Argument:

<arg> = {10e-3, 1e-3, 100e-6} seconds, or

{10ms, 1ms, 0.1ms}, or

{10000us, 1000us, 100us}

Valid Modes:

1 GHz Calibrator Preset Pulse control Mode

OUTPut:PULse:PERWID

Description:

Set or return the pulse period <arg1> and pulse width <arg2> in seconds in the Variable pulse
control mode of the internal 1 GHz calibrator. This parameter has no effect in the Preset pulse
control mode or the External pulse source mode.

Syntax:

OUTPut:PULse:PERWID <arg1> sp <arg2>

Argument:

<arg1> = 28e-6 to 131.07e-3 (Pulse period)

<arg2> = 7e-6 to 65.535e-3 (Pulse width)

Valid Modes:

Any

OUTPut:PULse:SOURce

Description:

Set or return the pulse modulation source for the internal 1 GHz calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator‘s rear-panel EXT
PULSE
input. Note this input requires TTL compatible signal levels.

Syntax:

OUTPut:PULse:SOURce <asc>

Argument:

<asc> = {INT, EXT}

Valid Modes:

Any

OUTPut:SIGNal

Description:

Set or return the on/off state of the internal 1 GHz calibrator‘s output signal.

Syntax:

OUTPut:SIGNal <asc>

Argument:

<asc> = {OFF, ON}

Valid Modes:

Any.