Read:interval:pkavg, Read:marker:average, Read:marker:delta – Boonton 4500B Peak Power Meter User Manual

Page 262: Read:marker:maximum, Read:interval:pkavg? -62, Read:marker:average? -62, Read:marker:delta? -62, Read:marker:maximum? -62, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-62


READ:INTERval:PKAVG?

Description:

For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.

Syntax:

READ[1|2|6|7]:INTERval:PKAVG?

Returns:

CC, peak-to-average ratio of power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:MARKer:AVERage?

Description:

For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

READ[1|2|3|4|5|6|7]:MARKer[1|2]:AVERage?

Returns:

CC, average power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.




READ:MARKer:DELTa?

Description:

For the specified channel return the difference between MK1 and MK2. The units will be the same
as marker units.

Syntax:

READ[1|2|3|4|5|6|7]:MARKer:DELTa?

Returns:

CC, (MK1 – MK2)

Where CC is the measurement condition code.

Valid Modes:

All

READ:MARKer:MAXimum?

Description:

For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:MARKer[1|2]:MAXimum?

Returns:

CC, maximum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

Advertising