Table 4-8c. statistics > term settings > submenu, Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual

Page 114

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-42

Table 4-8C. Statistics > Term Settings > Submenu

Menu Item

(Type)

Selections

Function

Term Time

Stat

Set terminal run time

(Numeric)

Discrete ranges:

Off, 1, 2, 5, 10, 30, 60,

Use any of the data entry controls to set a

120, 300, 600, 1800,

running time for statistical data collection.

3600 seconds

When the time expires, the measurement will
execute the Terminal Action. Select Off to
ignore running time.


TRIGger:CDF:TIMe {0 to 3600}

Term Count

Stat

Set terminal sample count

(Numeric)

Continuous range:

2.0 to 4,096.0

Use any of the data entry controls to set a

Megasamples

terminal sample count. When the count is
reached, execute the Terminal Action.

TRIGger:CDF:COUNt <2 to 4096> (M)


Term Action

Stat

Select the terminal action to execute

(Mult. Choice)

Stop, Flush, Decimate

Stop – when the terminal time or count is
reached, stop the measurement.

Flush – when the terminal time or count is
reached, clear the histograms and buffers
and continue with a new measurement.

Decimate - when the terminal time or count
is reached, divide each histogram bin count
and the total sample count by 2 and continue.

This will provide a continuously updating Fig. 4-26.
distribution function.

Statistics > Term Count

menu.

TRIGger:CDF:DECImate {ON, OFF}