Table 4-8b. statistics > menu, Table 4-8b. statistics > menu -39, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 111: Operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-39

Table 4-8B. Statistics > Menu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Presentation

Stat

Linear or logarithmic x-axis probability display

(Toggle)

Lin, Log

Select Lin to display percent probability on a linear scale. Select Log to

display percent probability on a logarithmic scale.


DISPlay:STATistical:PRESentation {LIN, LOG}


Horiz Scale

Stat

Select horizontal scale

(Numeric)

Lin: Discrete range

0.1% / Div to 10% / Div Use any of the data entry controls to select a horizontal range. The

10% / Div range displays the total scope of probability values from

0% to 100%. For the more sensitive ranges, also use the % Offset

function (see, below) to position the expanded scale so that the point

of interest is on-screen.

An invalid combination of Horiz Scale and Horiz Offset is not allowed.

The Offset will be forced to a valid value.


DISPlay:PCTSCALE <10, 5, 2, 1, 0.5, 0.2, 0.1> (%)


Log: Discrete range

Use any of the data entry controls to select the number of decades of

1 to 7 decades

probability displayed on the horizontal axis.

DISPlay:DECAdes <1 to 7> (decades)



Full Scale

Stat

Select the right edge Horiz Scale (Log) value

(Numeric)

Discrete decade ranges:
Log: 0.01 % to 100 %

For the logarithmic scale only, use any of the data entry controls to
select the percent value of the right hand edge of the display. Allows
the display to be shifted to the right by 4 additional decades to display
probability down to 1E-9 %. See Figure 4-25b.

DISPlay:FULLscale <.01 to 100> (%)