Boonton 4500B RF Peak Power Analyzer

Operation

4-60

Table 4-13. Ref > Menu (continued)

Menu Item

Function

(Type)

Selections

(SCPI cmd)


To remove Reference lines from the display, select Off. To make the
Reference line refer to the active channel, select Active. To refer to a
specific channel, select it from the list.


REFLine:ASSIgn {OFF, ACTIVE, CH1,

CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}



Ref Tracking

Pulse

Control reference line tracking

(Mult. Choice)

Off,
Markers,

In reference line tracking mode the instrument sets the level reference

Top-Bottom,

markers to the position indicated by the Ref Tracking menu description

Distal-Mesial,

on the reference channel selected.

Distal-Prox.,


When tracking is ―Off‖, the reference lines are set to the levels entered

Mod

in the Ref Line1 and Ref Line 2 menus. The ―Markers‖ tracking

Off, Markers

selection sets the reference lines to the marker levels. Marker 1 is
loaded into Ref 1 and Marker 2 is loaded into Ref 2. The ―Top –

Stat

Bottom‖ selection loads the automatic measurement of the Top line

Off

amplitude into Ref 1 and the Bottom line amplitude into Ref 2. The
Distal - Mesial selection loads the automatic measurement of the distal
level into Ref 1 and the Mesial level into Ref 2. The Distal – Proximal
selection loads the automatic measurement of the Distal into Ref 1 and
the Proximal level into Ref 2.

REFLine:TRACk {OFF, X1X2, TOPBOT, DISTMESIAL,

DISTPROX}



Ref to Mks

Pulse, Mod & Stat



Sets the levels of both of the Ref lines to the levels displayed by the

(Action)

Set measurement markers MK1 and MK2

.

Ref 1 is set to MK1 level, Ref2 is set to
MK2 level.

REFLine:POSItion:FORCetomk