Read:marker:minimum, Read:marker:ratio, Read:marker:rdelta – Boonton 4500B Peak Power Meter User Manual

Page 263: Read:marker:rratio, Read:marker:minimum? -63, Read:marker:ratio? -63, Read:marker:rdelta? -63, Read:marker:rratio? -63, Boonton 4500b rf peak power analyzer, Remote operation

Advertising
background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-63


READ:MARKer:MINimum?

Description:

For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:MARKer[1|2]:MINimum?

Returns:

CC, minimum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

READ:MARKer:RATio?

Description:

For the specified channel return the ratio of MK1 to MK2. The units will be dB for logarithmic
units or percent for linear units..

Syntax:

READ[1|2|3|4|5|6|7]:Marker:RATio?

Returns:

CC, MK1/MK2

Where CC is the measurement condition code.

Valid Modes:

All




READ:MARKer:RDELTa?

Description:

For the specified channel return the difference between MK2 and MK1. The units will be the same
as marker units.

Syntax:

READ[1|2|3|4|5|6|7]:MARKer:RDELTa?

Returns:

CC, (MK2-MK1)

Where CC is the measurement condition code.

Valid Modes:

All

READ:MARKer:RRATio?

Description:

For the specified channel return the ratio of MK2 to MK1. The units will be dB for logarithmic
units or percent for linear units..

Syntax:

READ[1|2|3|4|5|6|7]:MARKer:RRATio?

Returns:

CC, MK2/MK1

Where CC is the measurement condition code.

Valid Modes:

All