Table 4-9A. Trig > B Trig > Submenu (continued)

Menu Item

Function

(Type)

Selections

(SCPI cmd)



B Source

Pulse & Stat*

Same as Trig > Source for the A Trigger, Table 4-9.

TRIGger:BTRIGger:SOURce {CH1, TRIG1, CH2, TRIG2}



Trig Level

Pulse & Stat*

Same as Trig > Trig Level for the A Trigger, Table 4-9 .


TRIGger:BTRIGger:LEVel <-39.99 to 20.00> (dBm typ)



B Slope

Pulse & Stat*

Same as Trig > Trig Slope for the A Trigger, Table 4-9.


TRIGger:BTRIGger:SLOPe {POS, NEG}



B Dly Time

Pulse & Stat*

Set delay time for B Dly by Time mode

(Numeric)

Continuous range:
10 ns to 1 sec

Set the time delay that must occur after the A trigger before the sweep
starts. If the time delay is set to zero, the sweep will start on the first B
trigger following the A trigger (minimum delay). If the time delay is set
to T seconds, the sweep will start on the first B trigger that follows the
A trigger plus an interval of T seconds. If there is no B trigger, the
sweep will not start.


TRIGger:BTRIGger:TDLY <0 to 10000e-3> (sec)



B Dly EvntCnt

Pulse & Stat*

Set trigger event count for B Dly by Evnt mode

(Numeric)

Continuous range:
0 to 999,999

Set the number of B trigger events that must occur after the A trigger
before the sweep starts. When the count is set to zero, the sweep will
start when the A trigger occurs (no delay). If the count is set to 1, the
sweep will start on the first B trigger following the A trigger.). If the
count is set to N, the sweep will start on the Nth B trigger following the
A trigger.


TRIGger:BTRIGger:ECOUNT <0 to 999999>